Abstract

A method has been devised for testing ultraviolet monochromators for nearby stray radiation i.e., the stray radiation with wavelengths close to that of the dispersed radiation. The method is based on the observed transmittance of the very narrow 2537 A absorption peak of saturated mercury vapor at elevated temperatures. The usual method of determining stray radiation does not detect nearby stray radiation because the peak normally used for the measurement is so broad as to absorb this radiation. Nearby stray radiation is particularly objectionable in the recording of absorption spectra of the very sharp peaks observed at low temperature and in the spectra of vapors.

© 1955 Optical Society of America

Full Article  |  PDF Article
OSA Recommended Articles
Resolution and stray light in near infrared spectroscopy

Wilbur Kaye
Appl. Opt. 14(8) 1977-1986 (1975)

A Grating Spectrophotometer for the Ultraviolet and Visible Regions

T. B. Thomas and E. E. Schneider
J. Opt. Soc. Am. 41(12) 1002-1005 (1951)

Test Methods in Spectrophotometry: Stray-Light Determination

Richard E. Poulson
Appl. Opt. 3(1) 99-104 (1964)

References

You do not have subscription access to this journal. Citation lists with outbound citation links are available to subscribers only. You may subscribe either as an OSA member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access OSA Member Subscription

Cited By

You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an OSA member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access OSA Member Subscription

Figures (1)

You do not have subscription access to this journal. Figure files are available to subscribers only. You may subscribe either as an OSA member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access OSA Member Subscription